The chromatogram is the most important picture on a peptide certificate of analysis (COA). It shows how pure a sample is, and once you know what to look for, it takes about a minute to read.
What HPLC does
High-performance liquid chromatography (HPLC) pushes a dissolved sample through a column. Different molecules travel through at different speeds, so they come out at different times. A detector, usually measuring UV light absorbed at 214 or 220 nm where peptide bonds absorb, records a signal as each component leaves the column.
The two axes
- Horizontal axis (retention time): how long after injection each component came out, usually in minutes.
- Vertical axis (absorbance or signal): how much signal the detector saw. Taller and wider peaks mean more material.
What to look for
- One dominant main peak. A clean peptide sample shows one large peak that holds most of the total signal.
- Small side peaks. These are usually related impurities, such as truncated or modified peptides left over from synthesis. A few small ones are normal.
- A flat baseline. A smooth baseline means the signal is easy to measure. A noisy or drifting baseline makes the numbers less reliable.
- The early solvent peak. A peak right at the start often comes from the solvent, not the sample. A good report does not count it.
- The purity percentage. The report calculates the main peak’s area as a share of the total peak area. A result such as 99% means the main peak is about 99% of the detected signal.
What the chromatogram cannot tell you
- It does not confirm identity. A tall peak shows one dominant substance, not which substance it is. Mass spectrometry confirms the molecular weight.
- It is not net peptide content. Purity does not account for water or counterions in the powder. Read our guide to purity vs. net peptide content.
- It depends on the method. A different column, gradient or wavelength can show different peaks, so compare results only when the method is stated.
Red flags on a report
- No batch or lot number, or one that does not match your vial.
- No lab name or test date.
- A cropped chromatogram, with no peaks labeled or no method listed.
- A purity number with no chromatogram attached.
Browse lot-matched reports on our COA page, read about our testing standards, or start with the basics in how to read a peptide COA.
All products are sold for laboratory and research use only. Not for human or veterinary use.